Welcome to use our company's products - hollow glass dew point meter. The detector is based on our company.
GB/T11944-20
12 "Insulating Glass Standard" test dew point method developed. This equipment uses cascade refrigeration technology, intelligent digital display instrument, PID temperature control, with simple operation, intuitive display, fast cooling speed, accurate temperature control and so on.
Insulating glass dew point meter
Technical Parameters
1,
Power: 1000W
2,
Power supply requirements: AC 220V 50HZ
3,
Temperature control accuracy: ±1.0 °C
4,
Temperature resolution: 0.1 ° C
5,
Time resolution: 1s
6,
Minimum temperature: -70 ° C
7,
Working environment: 23 °C ± 2 ° C
Insulating glass dew point meter
About parameter setting
1,
Time parameter setting
Press >
â€
Key, time display, the lower row of the tube of the instrument is shining, press
â€
∧
â€
The key can be modified. (the meter unit time is seconds)
2,
Temperature parameter setting
Since all the parameter settings, PID setting and control parameters of the "temperature control instrument" of the equipment before leaving the factory are adjusted, do not change. Just press on the temperature control meter
"∧
â€
,
"
∨
â€
The “Set Temperature†(SV) can be modified by pressing the button.
3. For other parameter settings of “Timer Meter†and “Temperature Control Instrumentâ€, please refer to the instrument manual.
Insulating glass dew point meter
Instructions
1
Connect the power cord plug, turn on the “Power†switch on the control panel, power on the “Cooling Temperature†table and the “Test Temperature†table and the “Timer†table. The system is powered normally. The "test temperature" PV shows the current "cold probe" temperature.
2. Set the temperature of the "test temperature" SV according to the experimental standard, press "Start"
â€
Button, the indicator light is on, the system delay starts, and the work starts.
.
(Note: Because the equipment is a cascade refrigeration technology, this is the first-stage compression system startup work. After about 7 minutes, the secondary compression system starts automatically. The “temperature control instrument†PV on the control panel shows the cooling. Is the normal process requirements, don't worry!)
3. When the temperature reaches the “set temperatureâ€, start the experiment, press the chronograph button switch, and start timing according to the set time. After the time arrives, the alarm light will be on. Continue to the next experiment.
4. After the test is finished, press "Start"
â€
Button, the indicator light off the system stops working. Then turn off the power switch and turn off the power.
Insulating glass dew point meter
use
Precautions
1,
This equipment uses a compressor refrigeration process. It should be placed at rest for more than 4 hours before using the equipment for the first time. The time between starting the system twice should be no less than 30 minutes!
2,
When operating the equipment, be careful of frostbite due to the low temperature of the "cold probe"!
3,
Non-professional professionals are not allowed to disassemble the equipment. In order to avoid danger or damage to the equipment!